Features
- Solid robust welded steel framework housing the following:
- Gas panel constructed of porous refractory tiles in a stainless steel housing, mounted on a on the framework on a rotating steel ring
- Gas Panel can use natural gas, methane or propane mixed with air prior to entry to the panel
- Incorporates a dual gas safety shut off system
- Variable area calibrated flowmeters are incorporated in the supply pipework allowing gas and air supplies to be controlled and monitored
- Pressure governors, control and shut-off valves for the gas and air
- Safety flame arrestors on pilot burner system
- An adjustable pilot burner using acetylene/air mixture is used for the pilot flame
- A heat release stack is mounted on the specimen support framework directly above the test specimen complete with specially manufactured type K thermocouples and positioning frame
- Three specimen holders
- Three spring-loaded action clamp
- One dummy specimen
- Three specimen backing boards
- Calibration board of calcium silicate based material with a series of holes for positioning the heat flux meter at the prescribed calibration locations, together with plugs for the hole positions not in use during calibration
- Specially calibrated water-cooled Schmidt-Boelter heat flux meter with calibration certificate
- Thermocouple conditioning device for the heat release stack showing compensated signal and allowing connection to logger or recorder
Optional Extras
- Line burner - used to calibrate the heat release stack [relates to IMO Res A.653(16)]
- Air blower to provide an air supply at 9 litres/sec
- Software package and data acquisition system to record outputs (not including PC)
Services required
- Air at 9 litres/sec
- Gas - natural gas or methane supply
- Acetylene supply
- Electrical Supply - 230V/50Hz or 120V/60Hz